CtrlK
BlogDocsLog inGet started
Tessl Logo
SkillPublic

medical-device-software

Use for any question about the software inside a medical device, SaMD or regulated health app - which IEC 62304 safety class (A/B/C) it is and how much process and documentation that class needs, what to record about open-source or third-party libraries (SOUP, SBOM), trace matrix and release testing, IEC 81001-5-1 security, or the FDA documentation level and 524B cyber-device duties. Checks the user's files with four scripts that cite the clause.

Quality

94%

Does it follow best practices?

Run evals on this skill

Adds up to 20 points to the overall score

View guide
SecuritybySnyk

Passed

No findings from the security scan

SKILL.md
Quality
Evals
Security

Medical Device Software

Portability: 4 stdlib scripts, no network. Fill CSVs, run, keep the output with the software file. Not a conformity assessment. IEC 62304 is paywalled: clause markers here come from two agreeing secondary sources (🟡) and your licensed copy prevails. A notified body or FDA reviewer can read a borderline call differently.

Software files fail review in the same places: a class claimed lower than the harm allows, an "external" risk control that lives in the same software system, a SOUP list with "latest" as the version and no anomaly review, a trace matrix with risk controls nobody tested, and security treated as an afterthought. Three classifications get confused, and they use different inputs:

QuestionDecided byAssessed
IEC 62304 class A/B/C (which activities)worst-case harm, clause 4.3 + A1 Fig. 3after risk controls external to the software system
FDA Basic / Enhanced (what you submit)probable risk of death or serious injurybefore any risk controls
FD&C Act 524B cyber devicesoftware + any internet-capable path (USB counts)not risk-based

Workflow

# 1 - Classify: system and item classes, FDA level, cyber device?
python scripts/software_safety_classifier.py --inventory assets/software_inventory.example.csv
# 2 - Gap check: is every activity the class requires done, with evidence?
python scripts/lifecycle_gap_check.py --class B --security --assessment assets/lifecycle_assessment.example.csv
# 3 - SOUP: identified, specified, anomaly lists reviewed, in the SBOM?
python scripts/soup_register_check.py --register assets/soup_register.example.csv --sbom assets/sbom.example.cdx.json --cyber-device
# 4 - Release: every requirement tested, every risk control verified?
python scripts/trace_matrix_check.py --matrix assets/trace_matrix.example.csv --class C --release

Class first: steps 2 to 4 take the class from step 1. --template prints an empty input; lifecycle_gap_check.py --list --class X prints the required ids.

The gates block on (exit 4)

  1. A claimed class below what the harm allows, or an item below its system without a segregation rationale — 62304 4.3.
  2. An external risk control used to lower the class that is not itself verified — 62304 4.3 (A1).
  3. A required activity NOT MET, or NA where the clause is not conditional — the class table (A1 Table A.1).
  4. A SOUP item with no exact version (8.1.2), no stated requirements (5.3.3/5.3.4) or no anomaly-list review (7.1.3) for class B/C, or past end of support with no plan.
  5. A class B/C risk control with no hazard or no passing test (7.3.1, 7.3.3); with --release, any untested requirement or open failed result without an anomaly record (5.7, 5.8).
  6. A cyber device with no vulnerability monitoring on a SOUP item — 524B(b)(1).

Hard rules

  • Unassessed is not MET. The gap check refuses a file that skips a required id.
  • Unclassified is class C. Unknown inputs never round down.
  • Severity decides the class, not probability. Only verified external controls lower it.
  • Class A still tests. Software system testing (5.7) applies to every class since A1.
  • Mark the confidence. 🟢 primary text checked · 🟡 secondary sources agree · 🔴 not verified. EN 62304 and EN IEC 81001-5-1 are not confirmed as cited in the OJ under the MDR; never claim presumption of conformity without checking the current list.
  • Never sign or submit. The scripts prepare evidence; the manufacturer's named people approve the class, the release and the submission.

Scripts

ScriptRoleExit codes
scripts/software_safety_classifier.py62304 4.3 class per system and item, segregation, FDA documentation level, 524B cyber device0 classified · 2 findings · 3 bad input · 4 blocked
scripts/lifecycle_gap_check.py62304 activities by class (+4.4 legacy), 81001-5-1 activities with --security0 ready · 2 gaps · 3 · 4 blocked
scripts/soup_register_check.py8.1.2, 5.3.3/5.3.4, 7.1.3, end of support, 81001-5-1 risk-transfer category, SBOM cross-check (CycloneDX/SPDX)0 pass · 2 · 3 · 4 blocked
scripts/trace_matrix_check.pysystem → software requirement → test → result, risk control → hazard, release state0 pass · 2 · 3 · 4 blocked

All support --help, --sample, --template and --output json.

References and assets

  • references/classification_and_documentation_levels.md — 62304 class decision path, external controls, segregation, FDA Basic/Enhanced, 524B, EU status, Edition 2 (10 sources)
  • references/lifecycle_requirements_by_class.md — every clause by class, legacy software, regulatory standing (7 sources)
  • references/secure_lifecycle_81001_5_1.md — 81001-5-1 clause map, ISH1:2025 risk transfer, mapping to 524B and MDCG 2019-16 (7 sources)
  • references/soup_and_traceability.md — SOUP evidence by class, SBOM, what reviewers check in a trace matrix (8 sources)
  • assets/*.example.csv, assets/sbom.example.cdx.json — input formats, generated from each script's --sample
Repository

alirezarezvani/claude-skills

Path

ra-qm-team/medical-device-software/skills/medical-device-software/SKILL.md

Last updated
First committed

Also appears in

alirezarezvani/claude-skills
In sync

since Nov 11, 2026

Is this your skill?

If you maintain this skill, you can claim it as your own. Once claimed, you can manage eval scenarios, bundle related skills, attach documentation or rules, and ensure cross-agent compatibility.